Patient weight gain after change of acid concentrate supplier with same nominal dialysate sodium level. PDF Print
IJAO: Conclusions: Changing dialysate acid concentrates, both labeled 1:44 dilution, led to the delivery of a higher dialysate sodium, resulting in weight gains, increased pre-dialysis blood pressure, but less symptomatic intradialytic hypotension. Following readjustment of volumetric dialysate mixing, excess weight gains and increased blood pressure resolved over 4 weeks, highlighting the importance of checking the delivered dialysate sodium following a change in dialysate acid concentrate.

...

 